MRI implant screening: a step-by-step verification workflow
How imaging centers identify an implant, read its MR Conditional labeling, match it to their scanner and document the decision, before the patient is on the table.
Published By Yashaswini Lagadapati
MRI implant verification is the process of identifying a patient's implant by type, location, make and model, finding its current MR safety labeling, and confirming the scan can meet every condition on that label. MR safety staff document it, and a designated radiologist makes the final call, ideally before the day of the exam.
Watch it in 74 secondsTranscript
MRI implant screening, step by step.
MR Conditional devices each have their own rules.
Pacemakers usually need reprogramming and monitoring.
Neurostimulators often have tight heating limits.
Aneurysm clips need the exact model.
Ask at booking about pacemakers, stimulators, implants, metal in the eye, and pregnancy.
Any yes or not sure starts verification.
On the day, screen again before Zone IV.
The implant card gives the type, maker, and model.
Find its current label, match a scanner, and get the radiologist's approval.
In Gravity, the questions follow the patient.
On the call, by text, at the kiosk.
A yes stops self-booking, so staff review it first.
Prep texts remind patients to bring their card.
Answers steer which slots are offered, and an AI assistant finds the device in an operative note.
Agents work the order. Your people work the exceptions.
See Gravity in action. Book a demo.
MRI implant verification in one minute
What it is: the work of finding out exactly which implant a patient has, what its MR safety labeling allows, and whether your scanner and protocol can stay inside those limits.
Why it matters: the magnet is always on. Ferromagnetic implants can move, conductive ones can heat, and active ones, such as pacemakers and neurostimulators, can malfunction.
The four steps: identify the device, find its current labeling, match the conditions to your scanner and exam, and document the decision.
Who decides: the MR Medical Director or a designated Level 2 MR physician makes the final call. The MR Safety Officer runs the process day to day.
Where it leaks: the implant nobody asked about at booking, the card the patient left at home, the label checked against the wrong field strength, the patient sent home from the scanner room and the slot that sits empty.
What fixes it: ask at booking, verify before the day of the exam, book the patient on a scanner that meets the conditions, and screen again before Zone IV.
Why implant verification belongs before the day of the exam
Most centers screen every MRI patient on arrival, as they should. The ACR Manual on MR Safety asks for a written or electronic screening form for every conscious, non-emergency patient before they enter Zone III, and for each patient to be screened twice, at least once by Level 2 MR personnel. Then the technologist does a full stop and final check at the door of Zone IV. The Joint Commission, which accredits many imaging providers, also expects MRI screening to cover implants and devices.
The problem is timing. When the first time anyone hears about a spinal cord stimulator is at the screening desk, there is rarely time to find the model, read a 40-page manual and reach the radiologist. The patient is rebooked, the slot goes unused, and the referring office waits another week for the report. The safest workflow is also the cheapest: catch the implant at booking and verify it while there is still time.

When the implant is verified at booking, the patient reaches the scanner. When it surfaces on the day, the patient often goes home.
In Gravity: MRI safety questions are asked once, tied to the exam, and kept on the patient. The same questionnaire appears wherever the patient meets the center: when booking staff take the call, when the patient books from a text, during pre check-in on their phone, on the lobby kiosk, and on the AI voice agent's calls, which ask screening questions. Questions appear only for the patients they apply to. The answers steer which slots are offered, and an answer that needs a person can stop a patient from finishing a self-booking and ask them to call, so staff can review the case before they book it.
MR Safe, MR Conditional, MR Unsafe: reading the label
Every implant decision starts with one of three terms, used by the FDA and defined in ASTM F2503:
Term | What it means | What to do |
|---|---|---|
MR Safe | Nonmetallic, nonconducting and nonmagnetic. Safe in all MR environments | Document it. A nonmetallic hernia mesh is an example |
MR Conditional | Shown to be safe only when specific conditions are met | Find the current conditions for the exact model and confirm the scan can meet every one |
MR Unsafe | Presents a known risk in the MR environment | Don't scan, except through a documented risk-benefit decision by the responsible physician |
A device that contains metal can never be MR Safe, only MR Conditional or MR Unsafe. Most of the work is in MR Conditional, because the conditions are specific.

An MR Conditional label is a set of limits: field strength, the pull near the bore, RF heating, scan time and the patient's position. The scan goes ahead only when every one is met.
FDA's 2023 labeling guidance lists what MR Conditional labeling may include:
Condition | What it limits | What to check at your center |
|---|---|---|
Static field strength (B0) | Which magnets the device was tested in, usually 1.5T, 3T or both | Which of your scanners qualify. A 1.5T-only device can't go on the 3T |
Maximum spatial field gradient | The pull on the device as the patient moves into the bore, in T/m or gauss/cm | Your scanner's spatial field gradient map, from the scanner's manual, at the implant's position |
RF exposure | Heating, as whole-body or head SAR, or B1+rms, and the operating mode (Normal or First Level) | That every sequence in the protocol stays inside the limit, which may mean a modified protocol |
Gradient slew rate | How fast the gradients switch | Scanner and sequence settings |
Scan time | Minutes of continuous scanning, and any wait between scans | The exam length, and whether two exams in one visit fit the limit |
Position and exclusion zones | Where the landmark may be, and which body regions can't be at isocenter | Whether the exam ordered can be done at all |
Before and after the scan | Device programming, checks, or removal of external parts | Who does it, and whether it needs a cardiology or device representative |
Three cautions. Conditions apply only to the exact make and model tested, so "a stent" or "a hip replacement" isn't enough on its own. Conditions change, so use the most recent instructions for use, not the copy in the patient's folder from five years ago. And the conditions for the same device can differ from one country to another, so use the labeling for the US.
The workflow, step by step

Identify the device, find its current labeling, match the conditions to your scanner, and document the decision.
Ask at booking. Ask every MRI patient a short set of safety questions when the exam is booked: pacemaker or defibrillator, neurostimulator, cochlear implant, aneurysm clip, implanted pump or port, other implants or surgical hardware, metal in the eye, shrapnel or bullets, recent surgery, and pregnancy. Any "yes" or "not sure" starts a verification before the patient is given a slot.
Identify the device. You need its type, its location in the body, the manufacturer and the model, and ideally the implant date. The best sources are the patient's implant card, the operative note, the implant list in the medical record, and earlier MR screening forms. When the patient doesn't know, call the implanting surgeon's office. Record the identification in writing or electronically.
Find the current labeling. Go to the manufacturer's MRI safety information or instructions for use for that model, or call the manufacturer with the patient's details. Published reference lists help, but the manufacturer's current labeling wins when they disagree.
Match the conditions to your scanner and the exam. Check every condition against a specific scanner and a specific protocol: field strength, spatial field gradient at the implant's position, RF limits, scan time and landmark. This is where an MR Safety Expert earns their keep. Some exams turn out to be possible only on one of your magnets, or only with a modified protocol.
Book it to match. Book the patient on the scanner that meets the conditions, with any extra time the modified protocol needs, and with the device programming arranged if there is one. Send prep instructions that ask the patient to bring their implant card and, for some devices, their remote or charger.
Decide and document. The MR Medical Director or designated Level 2 MR physician approves the scan, or doesn't. Record the device, the labeling version used, the conditions, the scanner and who approved it, so the technologist on the day doesn't have to repeat the work.
Screen again on the day. The patient completes the full screening form, a Level 2 technologist reviews it, and the technologist does the full stop and final check before Zone IV, confirming the device and conditions match what was approved. External devices, such as insulin pumps, glucose monitors and hearing aids, come off and are stored.

The four MR safety zones. Screening happens before Zone III, and the technologist's full stop and final check happens at the door of Zone IV.
In Gravity: once the exam is booked, the prep instructions linked to it go to the patient on their own, so a reminder to bring the implant card reaches every patient booked for that exam. Before the visit, a text link lets the patient confirm their details, answer the screening questions, sign consents and upload documents from their phone, with no app and no password, and check-in can ask for photos of other documents the center needs, each one required or optional. Finishing check-in files the signed forms and answers on the order. An alert on the order is a hard stop that booking staff must acknowledge before they book, and every record shows who changed which field and when.
Who does what
Role | Part in implant verification |
|---|---|
Booking staff | Ask the safety questions, start verification on any "yes", book only once the case is cleared or a conditional slot is agreed |
MR Safety Officer (MRSO) | Runs the day-to-day program: gathers the device details and labeling, checks the conditions, prepares the case for the radiologist |
MR Safety Expert (MRSE) | Advises on the physics: spatial field gradients, RF heating, scanner-specific limits. Often a medical physicist |
MR Medical Director (MRMD) or Level 2 MR physician | Makes the final decision to scan, and the risk-benefit decision when the labeling can't be met or found |
MR technologist (Level 2) | Reviews the screening form with the patient, does the full stop and final check, runs the approved protocol |
Referring office | Supplies the operative note or device details, and documents medical necessity when a risk-benefit decision is needed |
Device classes that need the most care
The ACR manual lets a center decide, in its own written procedures, that some implant classes carry little MR risk, coronary stents and orthopedic hardware being common examples, and skip the exact make and model for them. Other classes always need exact identification:
Pacemakers, ICDs and other cardiac devices. MR Conditional status applies to the whole system, the generator and the leads together, and usually needs the device programmed into an MRI mode, a cardiology team available, and monitoring during the scan. Devices without MR labeling can be scanned under a program that follows the Heart Rhythm Society's 2017 consensus statement. Medicare has covered MRI for patients with pacemakers, ICDs and CRT devices, under conditions, since April 2018 (NCD 220.2). Many freestanding centers choose to refer these patients to a hospital, and should say so in writing and at booking.
Neurostimulators. Spinal cord, deep brain and vagus nerve stimulators often have narrow conditions: specific coils, low RF limits, specific body regions and a programming step before and after.
Cochlear implants and other active implants. Some need a head wrap or a magnet removed; some can't be scanned at all at a given field strength.
Intracranial aneurysm clips. The ACR manual says not to scan until the exact manufacturer, model and type is known, from documentation signed by a physician. Blanket letters saying all clips from a practice are safe aren't acceptable.
Implanted pumps and ports. Drug infusion pumps may need the pump checked after the scan; most ports are low risk but still need identifying.
External devices. Insulin pumps, continuous glucose monitors and hearing aids are removed before Zone IV.
When the implant can't be identified
Sometimes the patient has no card, the surgery was decades ago, or the manufacturer gives no MR labeling. The ACR manual describes a two-step process for these cases: first a risk assessment by the responsible MR personnel, then a risk-benefit decision on how to proceed, often involving the MR Medical Director, the MR Safety Officer, the MR Safety Expert and a radiologist. Write down what was known, what wasn't, who decided and why. When the scan goes ahead outside the labeling, the patient should understand the risk and consent to it, and the referring provider should document why the exam is needed.
A prior MRI isn't proof. The ACR manual is explicit: a patient scanned before without a known adverse event is not evidence that an unidentified implant is safe for the next scan. Field strength, the device's orientation and the way the patient moved through the field can all differ.
Metal in the eye. Any patient with a history of a potential ferromagnetic foreign body in the eye for which they sought medical attention needs the orbits evaluated before MRI: an orbit radiograph, or a radiologist's review of a thin-section CT taken since the injury. Earlier MR images alone don't clear the orbits.
Shrapnel, bullets and other foreign bodies. These need investigation before Zone III: history, plain radiographs, prior CT or recent MR of the area, and a judgment about the object's location and how long it has been there.
In Gravity: the case doesn't have to live in a hallway conversation. The order sits in a worklist where staff leave comments and alerts for each other, so everyone working the case sees the same notes, and staff can ask an AI assistant questions about the document open in front of them, such as finding the device named in a long operative note. The decision itself stays with the center's radiologist.
An implant found during the scan
Sometimes the first sign of an implant is on the images: a large susceptibility artifact where nobody expected metal. The ACR manual asks every center to have a written procedure for it. Stop scanning, keep the patient lying flat, bring the table out slowly, keep them recumbent until well away from the bore, and call the radiologist. Then file an incident report, so the screening question that missed it can be fixed.
A checklist you can adopt
Before the day | On the day |
|---|---|
Safety questions asked at booking | Full screening form completed by the patient |
Device type, location, make and model recorded | Form reviewed with the patient by a Level 2 technologist |
Current US labeling found and saved with the case | Device and conditions match what was approved |
Conditions checked against a named scanner and protocol | External devices removed, patient gowned |
Physician approval documented | Full stop and final check at the door of Zone IV |
Booked on the right scanner, with any extra time and device programming | Device checks after the scan, when the labeling asks for them |
Each row that slips to the day of the exam is a chance to send a prepared patient home. Each one done at booking is an exam that happens on the first visit.
It can run on top of the systems you already run. More than 460 imaging centers across the US trust Gravity, from the referral to the payment.
MR safety labeling and practice standards change. Confirm your center's implant policy with your MR Medical Director, your MR Safety Officer and your medical physicist, and always use the device manufacturer's current labeling.

Free download
MRI safety screening form
A three-page screening form you can adapt for your center: two pages for the patient, one for staff covering both screens and the implant decision.
Open the PDF →PDF, 3 pages, no sign-up
Get the editable kit
- The form in Word, to put your center's name, logo and policy on it
- The same questions as a Gravity questionnaire you can import
- The printable PDF
Sunrise Imaging is a fictional center. Adapt the form with your MR Medical Director before using it with patients.
Questions
What does MR Conditional mean?
MR Conditional means a device has been shown to be safe in the MR environment only when specific conditions are met. The labeling sets them: the static field strength, the maximum spatial field gradient, limits on RF energy such as SAR or B1+rms, scan time, and sometimes the patient's position, the coils or a device programming step. MR Safe items are nonmetallic, nonconducting and nonmagnetic, and are safe in any MR environment. MR Unsafe items should not enter it. A device that contains metal can only be MR Conditional or MR Unsafe.
Is an implant card enough to clear a patient for MRI?
An implant card is one of the best ways to identify the device, but it isn't a clearance on its own. The card tells you the manufacturer and model. The center still has to find the current MR conditions for that exact model, confirm its scanner and protocol can meet them, and document the decision. Operative notes, the medical record's implant list and earlier MR screening forms are also accepted sources.
Can a patient with a pacemaker or ICD have an MRI at an outpatient imaging center?
Sometimes. Many current pacemakers and ICDs are MR Conditional as a complete system, generator and leads together, and need the device reprogrammed into an MRI mode and the patient monitored during the scan. Devices without MR labeling can be scanned under a program that follows the Heart Rhythm Society's 2017 consensus statement, and Medicare has covered MRI for these patients, with conditions, since April 2018. Because a cardiology team has to program the device, many freestanding centers refer these patients to a hospital.
The patient had an MRI before with the same implant. Is that proof it's safe?
No. The ACR Manual on MR Safety says a prior scan without a known adverse event is not evidence that the implant is safe for future scans. The field strength, the device's position in the field and how the patient moved through it can all differ from one scan to the next. An earlier screening form can help identify the device, but clearance still rests on its labeling and a documented decision.
Who makes the final decision to scan a patient with an implant?
Under the ACR Manual on MR Safety, the final decision is made by the MR Medical Director or a designated Level 2 MR physician responsible for the patient, usually a radiologist. The MR Safety Officer runs the day-to-day work of finding and checking the labeling, and the MR Safety Expert, often a medical physicist, advises on technical questions such as spatial field gradients and RF heating.
How does Gravity help with MRI implant screening?
Gravity asks the center's MRI safety questions once, at booking, at pre check-in, on the lobby kiosk or by phone, and keeps the answers on the patient. The answers steer which slots are offered and can stop a patient from booking themselves and ask them to call, so staff hear about an implant before the day of the exam. The center's own staff and radiologists still identify the implant and make the safety decision.
Sources
- ACR Manual on MR Safety (2024): zones, MR personnel, screening, implant risk assessment and spatial field gradients
- US FDA: Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment (final guidance, October 2023)
- Indik JH, et al. 2017 HRS expert consensus statement on magnetic resonance imaging and radiation exposure in patients with cardiovascular implantable electronic devices. Heart Rhythm. 2017;14(7):e97-e153
- CMS National Coverage Determination 220.2: Magnetic Resonance Imaging (MRI for patients with pacemakers, ICDs and CRT devices)